Michell Easidew -100 to 20 Deg C 4-20 mA Output RS485 Communication Dew Point Sensor
The Easidew Transmitter is meticulously crafted for effortless usability, integrating all essential features to ensure hassle-free installation and operation. For the first time, dew point measurement is made as straightforward as temperature and pressure with this fully configured, calibrated transmitter, ready to be seamlessly integrated into your air or gas management and control system.
Highlights:
• 2-wire loop powered connection
• Dew point or ppm moisture content measurement
• IP66 (NEMA 4) protection rating
• Superior sensor protection
• Measurement range from -100 to +20°C (-148 to +68°F)
• Operating temperature range of -40 to +60°C (-40 to +140°F)
• Rapid response time
Performance:
Measurement range (dew point) -100 to +20°C (-148 to +68°F)
Accuracy (dew point) ±2°C (±3.6°F)
Response time: 5 minutes to T95 (dry to wet)
Repeatability: 0.5°C (0.9°F) dew point
Electrical output/input:
Output signal: 4-20 mA (2-wire connection, current source)
User-configurable over range
Output: Dew point, moisture content in ppmV
Output range: Dew point from -100 to +20°C (-148 to +68°F)
Moisture content in gas: 0-3000 ppmV
Supply voltage: 12-28 VDC
Load resistance: Max 250 Ω @ 12 V (500 Ω @ 24 V)
Current consumption: 20 mA max
Operating conditions:
Operating humidity: 0-100% RH
Operating temperature range: -40 to +60°C (-40 to +140°F)
Operating pressure: 45 MPa (450 barg / 6500 psig) max
Flow rate: 1 to 5 Nl/min (2.1 to 10.6 scfh) mounted in standard sampling block; 0 to 10 m/sec
standard sampling block; 0 to 10 m/sec
(0 to 32.8 fps) direct insertion
Temperature coefficient: Temperature compensated across operating temperature range